The Data Span/Set/Cal Menu Structure; Span Data - Siemens SITRANS FUS1020 Manual

Nema-4/ip 65 clamp-on transit-time flowmeter
Table of Contents

Advertisement

Section 6
Data Span/Set/Cal

Span Data

Set Alarm Levels
Calib. Flowrate
Use this menu to set the
low and high range for
analog outputs.
Use this menu cell to
enter a calibration
slope factor for
volumetric flow (Vfo).
6.7.1 SPAN DATA
The Span Data menu allows you to set 0% and 100% output limits for volumetric flow (Vfo), absolute
flow (Vfab) and sonic velocity (Vs). Each menu cell shows appropriate rate units and time base. If you
change flow rate units after spanning the system, the computer automatically updates the output data
setup to reflect the change. Span limits apply to both the analog outputs and the on-screen stripchart.
The flow outputs operate as follows:

THE DATA SPAN/SET/CAL MENU STRUCTURE

PGEN
Max Flow (Units)
Min Flow (Units)
Max Vs m/s
Min Vs m/s
Hi Flow
Low Flow
Interface Vs m/s
Aeration %
Makeup Latch
Intrinsic
Kc
MultiPoint
Siemens
Define data Alarm thresholds
Span Data
Set Alarm Levels
Calib. Flowrate
Data Span/Set/Cal
x.xxx (numeric entry)
x.xxx (numeric entry)
x.xxx (numeric entry)
x.xxx (numeric entry)
x.xxx (numeric entry)
x.xxx (numeric entry)
x.xxx (numeric entry)
x.xxx (numeric entry)
x.xxx (numeric entry)
Off
On
x.xxx (numeric entry)
Point #
Indicated Rate
% Correction
2 Channel (1)
Channel 1
Intrinsic
6-39
1020NFM-1D
xx (numeric entry)
x.xxx (numeric entry)
x.xxx (numeric entry)
Use this menu to define
setpoints for the alarm
relay outputs.