Delete Pipe; Transducer Type Menu - Siemens SITRANS FUS1020 Manual

Nema-4/ip 65 clamp-on transit-time flowmeter
Table of Contents

Advertisement

Section 6
To use a new name, press the <Right Arrow> and then use the appropriate PC keys or KeyPad
to rename the pipe (8 chars. Max.).
Press <ENTER>. This adds the new pipe to the pipe class.

Delete Pipe

This menu allows you to remove any of the pre-loaded pipes within a class from a pipe table. We
recommend that you use this function only to delete pipes that you have added so as to preserve the
factory presets.
To delete a pipe from the Pipe Table
To access the [Delete Pipe] menu press the <Right Arrow>. To access the [Choose Pipe
Class] option list press the <Right Arrow>.
Press the <Up/Down Arrow> scroll the option list to the class that contains the pipe to be
deleted.
To select the pipe class press <ENTER>. This moves the highlight to [Choose Pipe Name].
To access the [Choose Pipe Name] option list press the <Right Arrow>.
Press the <Up/Down Arrow> to scroll the option list to the pipe to be deleted.
To remove it from the pipe table press <ENTER>.

6.10.3 TRANSDUCER TYPE MENU

During the Transducer Install procedure, System 1020 analyzes the entered pipe and liquid data, then
automatically generates a list of the transducer sizes most suited for the application. The Transducer
Type menu allows you to place "marks" (asterisks) next to any transducers that you may want the
meter to consider preferentially during its recommendation routine. These transducers will be included
on the recommended list, that appears on the reverse-video prompt line at the top of the display screen.
The left-most transducer on the list is the most applicable while the right-most transducer on the list is
the least applicable. See the Transducer Type menu structure on the next page.
To mark Transducers:
To access the [Transducer Type] menu cell press the <Right Arrow>.
Press the <Up/Down Arrow> to scroll the highlight to a transducer type (e.g., 1011H High Pre-
cision) the press the <Right Arrow>to access the size option list.
Scroll through the option list by pressing the <Up/Down Arrow> and press <ENTER> to "mark"
desired transducers. Note that a plus sign (+) appears before each "marked" transducer.
If you want to clear a "marked" transducer, move the highlight to the transducer then press the
<ENTER> key. Note that this removes the plus sign.
To leave an option list press the <Left Arrow>.
6-57
1020NFM-1D