Upon completion, the meter beeps and the LCD screen shows the prompt:
You must now remove and then remount the upstream and downstream transducers in their re-
versed positions (i.e., Mount the Up Transducer with its cable in the Down Transducer/cable loca-
tion. Mount Down Transducer with its cable in the Up Transducer/cable location). If possible, this
procedure should be done without removing the transducer cables from the transducers. When
remounting the transducers, couple them to the pipe properly.
After re-installing the transducers, press [ENTER]. The computer measures the negative flow rate
briefly, then beeps and repeats the prompt: Reverse Xdcrs
mount the transducers for normal operation (in their original orientation). When remounting trans-
ducers, couple them to the pipe properly.
Siemens
2 Channel
ReversaMatic Active
Transducer Model
Transducer Size
Xdcr Mount Mode
Spacing Offset
Number Index
Spacing Method
Ltn Value
Install Complete?
Empty Pipe Set
Zero Flow Adjust
Pick/Install Xdcr
Reverse XDCRs
Siemens
2 Channel
Reverse XDCRs
Transducer Model
Transducer Size
Xdcr Mount Mode
Spacing Offset
Number Index
Spacing Method
Ltn Value
Install Complete?
Empty Pipe Set
Zero Flow Adjust
Pick/Install Xdcr
2-35
[1]
SITE1
[6:
:0]
1021 Xdcr
C1
Reflect
Invalidlow
4
Invalid
Auto Zero
ZeroMatic
ZeroClr
Actual Zero
>ReverseMatic
Press [ENT].
Press [ENT]. You must now re-
[1]
SITE1
Press [ENT]
1021 Xdcr
C1
Reflect
Invalidlow
4
Invalid
Auto Zero
ZeroMatic
ZeroClr
Actual Zero
>ReverseMatic
1020NFM-1D