Section 7
1020NFM-1D
Press the <Right Arrow> and then the <Down Arrow> to switch the option list to [Clr Active
Memory?
Yes]. Press <ENTER> to clear all Active Site Data (but not saved Site Setups).
To restore operation, type: MENU and press <ENTER> to access the Installation Menu. Create
a new Site Setup or recall a stored Site Setup. Re-select any Meter Facilities items (e.g., RS-
232 setup parameters).
To clear ALL Dynamic Memory using the RS-232 Interface:
Turn off power (if it is currently on).
Turn power on. As soon as you apply power type the @ character three times. The prompt: [Clr
Saved Data? Yes/No] appears at the top of the screen. Press the <Down Arrow>. Note that
the prompt switches to [Clr Saved Data?
No].
Press the <Right Arrow> and then the <Down Arrow> to switch the option list to [Clr Saved
Data?
Yes]. Press <ENTER> to clear all Dynamic Memory.
To restore operation, type: MENU and press <ENTER> to access the installation menu. Create
a new site setup and complete the installation procedure. Re-select desired Meter Facilities
items (e.g., RS-232 setup parameters).
7-15