To enter the Pipe Wall Thickness:
To enable numeric entry press the<Right Arrow>.
Use the PC numeric keys or KeyPad to enter exact wall thickness (use in/mm).
To register the pipe wall thickness press <ENTER>.

6.2.6 LINER MATERIAL

For lined pipes, select a pipe liner from the material option list. If the pipe liner material does not appear
on the list then select the closest type available. If necessary, call Customer Service for additional
help. The system default Liner Material is [None]. If the pipe does not have a liner, press the <Down
Arrow> twice to bypass to the next two menu cells. The [Liner Material] option list offers a selection of
common liner materials.
To select a Liner Material:
To access the [Liner Material] option list press <Right Arrow>.
To scroll to the required Liner material press the <Up/Down Arrow>.
To register selection press <ENTER>.

6.2.7 LINER THICKNESS

If you specified a pipe liner in the [Liner Material] menu cell, then use this menu cell to set its exact
thickness in appropriate units (English or Metric).
To enter a Liner Thickness:
To enable numeric entry press the <Right Arrow>.
Use PC numeric keys or KeyPad to enter the exact line thickness value.
To register the data press <ENTER>.
6.3

THE APPLICATION DATA MENU

This menu becomes available after picking a Meter Type, Measurement Channel and Measurement
Technology. Use of this menu is optional.
CAUTION: An erroneous Viscosity entry could affect the meter's intrinsic calibration. Do not
alter the default viscosity value of your liquid unless you are sure of your data.
Use the [Application Data] menu to edit default settings to match your application. When you specify a
Liquid Class, the meter will adjust its operation to accommodate the liquid's estimated sonic velocity,
viscosity and specific gravity. If necessary, you can edit each liquid class parameter individually to
obtain a closer match with the liquid. The default liquid is Water at 68
°
6-10
1020NFM-1D
°
F (20
C).