The Test Facilities Menu; Test Facilities Commands; Makeup - Siemens SITRANS FUS1020 Manual

Nema-4/ip 65 clamp-on transit-time flowmeter
Table of Contents

Advertisement

Section 7
7.6

THE TEST FACILITIES MENU

The Test Facilities menu provides commands for system analysis and recovery. The most useful for
the end-user is [Makeup] command. Using these routines under the supervision of our technical sup-
port staff will help us to provide technical analysis and solutions.
To start a Test Routine:
To scroll the cursor to the test routine press the <Up/Down Arrow>.
To access the test option list press the <Right Arrow>. Press <Up/Down Arrow> to move the
asterisk to [Yes].
To start the routine press <ENTER>.
To end the routine press the <Left Arrow>.

Makeup

Tx Up
Tx Dn
Fixed ALC
TX Up Fixed ALC
TX Dn Fixed ALC
* For technical service diagnostics.
Makeup
The Test Facility Makeup routine allows you to command the meter to re-acquire the operating param-
eters established during the Initial Makeup routine invoked by the [Install Completed?] command (see
paragraph 6.4 Pick/Install Xdcr).
To invoke the Makeup routine:
Press <Up/Down Arrow> to scroll the cursor to [Makeup].
Press the <Right Arrow> to access the test option list. Press <Up/Down Arrow> to move the
cursor to [Yes].
Press <ENTER> to start the makeup routine.
Siemens
2 Channel
Right arrow invokes an InProcess Makeup
Makeup
Tx Up
Tx Dn
Fixed ALC
Tx Up Fixed ALC
Tx Dn Fixed ALC
Test Facilities

TEST FACILITIES COMMANDS

Commands the meter to re-acquire the receive signal.
Forces upstream transmission only.*
Forces downstream transmission only.*
Select "Yes" to command amplifier to hold a constant gain.*
Fixed ALC while transmitting up only.*
Fixed ALC while transmitting down only.*
7-10
[1]
Channel 1
No
No
No
No
No
No
1020NFM-1D