Section 2
Press the <Left Arrow> to accept the Vs setting. The highlight moves to [Empty Pipe Set]. This
action indicates that the meter has accepted the new Vs setting and is now measuring flow.

Setting The Zero Flow Level

The [Zero Flow Adjust] menu cell becomes active after the meter completes the transducer install
procedure. When you invoke the Transducer Install, the meter performs an automatic routine (Auto
Zero) to establish a standard zero flow setting. However, AutoZero only supports the Reflect Mode
transducer mounting. It will be suppressed for a Direct Mode installation (like the example described in
this section). In addition, some conditions disallow the proper functioning of AutoZero even in Reflect
Mode. In such a case, the meter provides two alternate means of achieving a zero setting:
Actual Zero: requires zero flow in pipe.
ReversaMatic: obtain a zero by Xdcr reversal.
Zero Flow Adjust helps the meter to remove a possible zero offset due to pipe or transducer asymme-
try. Setting the zero flow level allows the meter to sense absolute flow direction and magnitude under
low flow conditions. Zero Flow Adjust also provides a ZeroClr routine to remove a zero offset setting
(done automatically or manually). In most cases, AutoZero works well and is desirable since it requires
no user input. When AutoZero is not possible, we recommend using the Actual Zero command. Actual
Zero requires you to stop flow to allow the meter to sense any zero offset. If not possible due to
application conditions, then the second alternative, ReversaMatic, allows zero level sensing when
liquid flow is at any velocity. However, you should use Actual Zero at the first opportunity. The ZeroClr
command instantly resets the zero flow registers to [0.000].

Performing The Actual Zero Routine

NOTE: Please reduce flow through the pipe to zero.
This example continues the sample application. We have successfully installed our C1H transducers
(in Direct Mode) and can stop flow for the purpose of performing an Actual Zero (see next page).
Siemens
2 Channel
Drive 53
0
36
Transducer Model
Transducer Size
Xdcr Mount Mode
Spacing Offset
Number Index
Spacing Method
Ltn Value (in)
Install Completed?
Empty Pipe Set
Zero Flow Adjust
Pick/Install Xdcr
2-32
[1]
SITE1
[6:
:0]
1021 Xdcr
C1
Reflect
InvalidFlow
4
Invalid
2.330
Install
Chan Not Setup
Chan Not Setup
1020NFM-1D