Direct Mode Transducer Installation; Dual Mounting Track Assembly/Transducer Installation; Connect Transducer Cables - Siemens SITRANS FUS1020 Manual

Nema-4/ip 65 clamp-on transit-time flowmeter
Table of Contents

Advertisement

Section 3
Integrated Flowmeter/Mounting Track Cover
Assembly secured to Mounting Track
3.5

DIRECT MODE TRANSDUCER INSTALLATION

Dual Mounting Track Assembly frames allow the mounting of transducers in the Direct Mode. The
combination of mounting frames and spacing guides is the recommended way to mount Direct Mode
transducers. The mounting frame establishes the axial alignment of the transducers, and allows you to
remove and replace either transducer while preserving their exact mounting location.

3.5.1 DUAL MOUNTING TRACK ASSEMBLY/TRANSDUCER INSTALLATION

NOTE: The following transducer and mounting track installation procedure applies to the 1020
Single Channel and 1020 Dual Channel Remote wall mounted flowmeters.
Locate wall position to mount 1020 Flowmeter.
Mount flowmeter and connect power cables as required (see drawing 1020-7).
Apply power to flowmeter.
Perform all required Installation Menu steps until the flowmeter issues the Number Index. Record
the Number Index value displayed in the [Pick/Install] menu. This value will be used to position
the transducers.
Number Index
Current Menu
Selection
Figure 9
Number Index 25
>Pick/Install Xdcr
3-19
1020NFM-1D
Figure 10

Connect Transducer Cables

W ide B eam
Wide Beam
C lamp-O n
Clamp-On
Ultrasonic
U ltrasonic
Flowmeter
Flow meter