Pipe Temperature Range; Pipe Configuration - Siemens SITRANS FUS1020 Manual

Nema-4/ip 65 clamp-on transit-time flowmeter
Table of Contents

Advertisement

Section 6
To edit the Density SG setting:
To activate numeric entry press the <Right Arrow>.
Use PC numeric keys or KeyPad to enter the Density SG value.
To register the data press <ENTER>

6.3.2 PIPE TEMPERATURE RANGE

Controlotron rates the 1021 and 1011 Universal transducers for operation up to 250°F. System 1010
series flowmeters should be used for higher temperature applications that exceed 250°F at transducer
mounting locations. Please refer to Section 4 for flow tube pressure and temperature ratings.

6.3.3 PIPE CONFIGURATION

The [Pipe Configuration] menu cell in the Application Data menu presents a list of descriptions of piping
configurations that could affect the flow profile characteristics (such as "Single Elbow"). Examine the
option list. Selecting a piping configuration that closely approximates conditions at or near your mount-
ing location allows the flow computer to compensate for the effect of upstream piping on flow profile.
The number of diameters between the upstream configurations and the transducer installation can be
numerically entered via the [Anomaly Diams] menu cell.
To select a Piping Configuration:
To access the option list press the <Right Arrow>.
To highlight the desired piping configuration press the <Up/Down Arrows>.
To register selection press <ENTER>.
The default pipe configuration (Fully Developed) will cause the flowmeter to use the conventional Rey-
nolds Compensation Table when compensating for liquid flow profile behavior.
Use this menu cell to enter the
number of pipe diameters
between the upstream
configuration and the
transducer installation.
Siemens
Dual Path [1]
Disignate Pipe Shape Near Transducers
Liquid Class
UniMass Table
Temperature Range
Pipe Configuration
Anomaly Diams
Application Data
6-14
Path 1
Water 20C/68F
Create/Edit Table
-40F to 250F
Fully Developed
1 Elbow
Dbl Elbow +
Dbl Elbow -
Valve
Expander
Reducer
Norm Entry
Header Inlet
Intrusions
1020NFM-1D
Use this menu cell to select
the pipe configuration that
most accurately represents
the upstream pipe condition.