Siemens SITRANS FUS1020 Manual page 76

Nema-4/ip 65 clamp-on transit-time flowmeter
Table of Contents

Advertisement

Section 3
Feed cable through cable bushing access hole, connect to transducer and press cable bushing
into hole (Figure 3).
Place supplied O-Ring gasket around the recess in the transducer's footprint (Figure 3). Apply a
1/4" bead of couplant to emitting surface of transducer.
Transducer
Cable
Cable
Bushing
Install transducer back end first and mate with pipe (Figure 4). Butt front of transducer against
track assembly and secure by tightening the transducer clamp.
NOTE: Transducer must be in contact with the mounting track reference stop (head of screw)
for accurate transducer positioning.
Transducer
Clamp
Seat O-Ring gasket to the pipe by pushing it off the corners of the transducer (Figure 5).
Install Mounting Track Assembly cover and secure with four screws.
Figure 3
Transducer
Figure 4
3-21
1020NFM-1D
O-Ring
Coupling
Compound
Front End of
Transducer