The Test Facilities Menu; Test Facilities Commands; Makeup - Siemens SITRANS F Operating Instructions Manual

Ultrasonic flowmeters portable flowmeter
Hide thumbs Also See for SITRANS F:
Table of Contents

Advertisement

Section 2

2.11.6 THE TEST FACILITIES MENU

The Test Facilities menu provides commands for system analysis and recovery. The most useful for
the end-user are [Makeup] and [Graph]. Using these routines under the supervision of our technical
support staff will help us to provide technical analysis and solutions.
To start a Test Routine:
To scroll the cursor to the test routine press
To access the test option list press
To start the routine press
To end the routine press

Makeup

Graph
Tx Up
Tx Dn
Fixed ALC
TX Up Fixed ALC
TX Dn. Fixed ALC
Graph AutoZero
* For technical service diagnostics.
Makeup
The Test Facility Makeup routine allows you to command the meter to re-acquire the operating param-
eters established during the Initial Makeup routine invoked by the [Install Completed?] command (see
paragraph 2.4 Pick/Install Xdcr Menu).
To invoke the Makeup routine:
Press
Press
to access the test option list. Press
ENT
Press
to start the makeup routine.
Siemens
2 Channel
Right arrow invokes an InProcess Makeup
Makeup
Graph
Tx Up
Tx Dn
Fixed ALC
Tx Up Fixed ALC
Tx Dn Fixed ALC
Graph AutoZero
Test Facilities
ENT

TEST FACILITIES COMMANDS

Commands the meter to re-acquire the receive signal.
Enables an on-screen marker display. - Cor Q should be greater
than 1.5. Larger number is better. Above 3 is desirable. SIG Q
should be less than 10. A value of 8 or less is desirable.
Forces upstream transmission only.*
Forces downstream transmission only.*
Select 'Yes' to command amplifier to hold a constant gain.*
Fixed ALC while transmitting up only.*
Fixed ALC while transmitting down only.*
Use to establish a standard zero flow setting.
to scroll the cursor to [Makeup].
[1]
Channel 1
No
No
No
No
No
No
No
No
. Press
to move the asterisk to [Yes].
to move the cursor to [Yes].
2-66
1010PFM-3G

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sitrans 1010aerwdp

Table of Contents