Operating Conditions; Process Conditions; Process Fluid Temperature Range; Density - Siemens SITRANS FC720 Operating Instructions Manual

Coriolis c
Table of Contents

Advertisement

Process conditions

9 Operating conditions

9.1 Process conditions
The pressure and temperature ratings presented in this section represent the
design values for the devices. For individual applications (e.g. marine
applications with option S2x) further limitations may apply according to the
respective applicable regulations. For details see chapter Application and
industry related standards [} 172] under the heading Marine approvals.
In this chapter, all values related to pressure are gauge pressure values.

9.1.1 Process fluid temperature range

Allowed process fluid and ambient temperature ranges in hazardous areas
depend on classifications defined by applications, refer to Temperature
specification in hazardous areas [} 46].
For SITRANS FC7x0 the following process fluid temperature ranges are available:
Temperature
range
Standard
Medium
High

9.1.2 Density

Meter size
FCS700 DN100
FCS700 DN150
FCS700 DN200
Density measuring range for NTEP custody transfer approval
Tab. 5: Density measuring ranges (C17)
Option
C17
7 M E 4 4 x
-
7
5 6
Order code
position 12
1, 4
2, 5
3, 6
-
8 9 10 11
12
13 14 15 16
Process fluid
Design type
temperature
in °C
(°F)
-50 – 150
Compact type
(-58 – 302)
-70 – 150
(-94 – 302)
-70 – 230
Remote type
(-94 – 446)
0 – 350
(32 – 662)
Measuring range of density
in kg/l (lb/ft³)
0 – 2 (0 – 125)
Measuring range of density
in kg/l (lb/ft³)
0,79 – 0,93 (49 – 58)
A5E52748387-AA, 2023/09
FC7x0
Operating conditions
-
Z
Order code
position 14
A, B
C, D, E, F, G, H
D, F, H
D, F, H
37 / 180