MEASure Subsystem
Example
MEAS:DIG:DATA2:BIT4? reads port 2, bit 4 (data line D2-4).
MEAS:DIG:DATA0:LW96:BIT95? reads port 0, bit 95 (data line D0-7).

MEASure:DIGital:DATAn[:type]:TRACe

MEASure:DIGital:DATAn[:BYTE]:TRACe <name> reads 8-bit port n after the
completion of the handshake and stores the data block in <name>.
MEASure:DIGital:DATAn:WORD:TRACe <name> reads 16-bit port n after the
completion of the handshake and stores the data block in <name>.
MEASure:DIGital:DATAn:LWORd:TRACe <name> reads the 32-bit port n after the
completion of the handshake and stores the data block in <name>.
MEASure:DIGital:DATAn:LW64:TRACe <name> reads the 64-bit port after the
completion of the handshake and stores the data block in <name>.
MEASure:DIGital:DATAn:LW96:TRACe <name> reads the 96-bit port after the
completion of the handshake and stores the data block in <name>.
Parameters
Comments
Keysight E1458A 96-Channel Digital I/O Module User Guide
*RST Condition: Set to input on all ports.
Parameter
Parameter Type
Name
DATAn
Numeric
<name>
String
:TRACe <name> is the keyword (maximum 12 characters) that specifies
the block where the data should be stored. This block must have been
previously defined by the [SOURce]:DIGital:TRACe:DEFine command.
This command will completely fill the named block. The defined block size
sets the amount of data read. The block size must by an integer multiple of
the [:type] keyword used in this command. For example, valid block sizes
for LWORd are 4, 8, 12, 16 etc.
Keysight E1458A Command Reference
Range of Values
BYTE none, 0 through 11
WORD none, 0, 2, 4, 6, 8, or 10
LWORd none, 0, 4, or 8
LW64 none or 0
LW96 none or 0
previously defined block name
(max 12 characters)
Default
0
None
69