Keysight E1458A Command Reference
Example
DIG:HAND3:DEL .005 sets the delay between the data output and the assertion
of the control line to true on 8-bit port 3 to 5 ms.

[SOURce:]DIGital:HANDshaken:DELay?

[SOURce:]DIGital:HANDshaken:DELay? queries for the time between data valid
and the assertion of the control line to TRUE. This form of the command operates
only on 8-bit ports and returns a decimal value between 0 and 0.015.
Parameters
Comments
Example
DIG:HAND0:DEL? queries the delay time between data valid and the assertion of
the control line to TRUE on 8-bit port 0.
98
Parameter
Parameter
Name
Type
HANDshaken
Numeric
Discrete
– MIN or DEF returns 0.000002. MAX returns 0.015.
– The delay time must be set to the same value on all ports used in a
multiple-port operation.
– :HANDshaken is the keyword used for commands relating to data
handshaking at 8-bit port n. The port number n must be the last character
of the keyword without spaces.
Range of Values
None, 0 through 11
MIN | MAX | DEF
Keysight E1458A 96-Channel Digital I/O Module User Guide