Keysight E1458A Command Reference
[SOURce:]DIGital:DATAn[:type]:MONitor?
[SOURce:]DIGital:DATAn[:BYTE]:MONitor? returns a decimal value indicating the
current state of the data lines on 8-bit port n. The value returned is in the range
of 0 to 255.
[SOURce:]DIGital:DATAn:WORD:MONitor? returns a decimal value indicating the
current state of the data lines on 16-bit port n. The value returned is in the range
of -32768 to 32767.
[SOURce:]DIGital:DATAn:LWORd:MONitor? returns a decimal value indicating
the current state of the data lines on 16-bit port n. The value returned is in the
range of -231 to (231 - 1).
[SOURce:]DIGital:DATAn:LW64:MONitor? returns two decimal values indicating
the current state of the data lines on 64-bit port n. The values returned are each
in the range of -231 to (231 - 1).
[SOURce:]DIGital:DATAn:LW96:MONitor? returns three decimal values
indicating the current state of the data lines on 96-bit port n. The values returned
are each in the range of -231 to (231 - 1).
Parameters
Comments
90
Parameter Name
Parameter Type
DATAn
Numeric
–
This command performs a readback of the actual condition of the selected
data port. The DIG:DATAn[:type][:VALue]? command returns the
programmed state of the data port.
–
The value returned is not affected by the programmed data line polarity.
The returned value indicates the actual TTL state of the data lines.
–
DIG:DATAn[:type]:MON? is the sequence used for commands relating to
data lines at port n. The port number n must be the last character of the
keyword without spaces.
–
The keyword LW32 may be used instead of LWORd.
–
Related Commands: [SOURce:]DIGital:DATAn[:type]:POLarity
–
*RST Condition: Mode is NONE on all ports.
Range of Values
BYTE none, 0 through 11
WORD none, 0, 2, 4, 6, 8, or 10
LWORd none, 0, 4, or 8
LW64 none or 0
LW96 none or 0
Keysight E1458A 96-Channel Digital I/O Module User Guide
[SOURce:] Subsystem
Default
0