Keysight E1458A Command Reference
MEASure:DIGital:DATAn[:type]:BITm?
MEASure:DIGital:DATAn:BYTE:BITm? reads the state of bit m of 8-bit port n after the
completion of the handshake.
MEASure:DIGital:DATAn:WORD:BITm? reads the state of bit m of 16-bit port n after
the completion of the handshake.
MEASure:DIGital:DATAn:LWORd:BITm? reads the state of bit m of the 32-bit port n
after the completion of the handshake.
MEASure:DIGital:DATAn:LW64:BITm? reads the state of bit m of the 64-bit port after
the completion of the handshake.
MEASure:DIGital:DATAn:LW96:BITm? reads the state of bit m of the 96-bit port after
the completion of the handshake.
Parameters
Comments
68
Parameter
Parameter Type
Name
DATAn
Numeric
BITm
Numeric
–
Input data is always assumed to be in binary format since only a single bit
of data is being read. The command returns either a 0 or 1.
–
The keyword LW32 may be used instead of LWORd.
–
:DATAn is the keyword used for commands relating to the data at port n.
The port number n must be the last character of the keyword without
spaces.
–
:BITm is the keyword that specifies the bit read by this command. Like the
DATAn keyword, no space can be between the keyword BIT and the m
parameter.
–
If n is omitted, port 0 is used. If m is omitted, bit 0 is used.
–
Related Commands: SOURce:DIGital:DATAn:POLarity
Range of Values
BYTE none, 0 through 11
WORD none, 0, 2, 4, 6, 8, or 10
LWORd none, 0, 4, or 8
LW64 none or 0
LW96 none or 0
BYTE 0-7
WORD 0-15
LWORd 0-31
LW64 0-63
LW96 0-95
Keysight E1458A 96-Channel Digital I/O Module User Guide
MEASure Subsystem
Default
0
0