[SOURce:] Subsystem
[SOURce:]DIGital:DATAn:LW96:TRACe <name> writes the named block of data to
the 96-bit port whenever the port is ready to start a new handshake.
Parameters
Comments
Example
DIG:DATA2:TRAC:WORD first_block writes data from the user memory block
first_block to 16-bit port 2.

[SOURce:]DIGital:DATAn[:type][:VALue]

[SOURce:]DIGital:DATAn[:BYTE][:VALue] [<base>]<value> writes data to 8-bit
port n. Values can be binary, octal, decimal, or hexadecimal.
[SOURce:]DIGital:DATAn:WORD[:VALue] [<base>]<value> writes data to 16-bit
port n. Values can be binary, octal, decimal, or hexadecimal.
[SOURce:]DIGital:DATAn:LWORd[:VALue] [<base>]<value> writes data to the
32-bit port n. Values can be binary, octal, decimal, or hexadecimal.
[SOURce:]DIGital:DATAn:LW64[:VALue] [<base>]<value> writes data to the 64-bit
port. Values can be binary, octal, decimal, or hexadecimal.
[SOURce:]DIGital:DATAn:LW96[:VALue] [<base>]<value> writes data to the 96-bit
port. Values can be binary, octal, decimal, or hexadecimal.
Keysight E1458A 96-Channel Digital I/O Module User Guide
Parameter Name
Parameter Type
DATAn
Numeric
<name>
String
The keyword LW32 may be used instead of LWORd.
:DATAn and :TRACe are the keywords used to write data to port n from
block <name>. The port number n must be the last character of the
keyword without spaces.
Related Commands: [SOURce:]DIGital:DATAn[:VALue]
[SOURce:]DIGital:DATAn:POLarity
*RST Condition: All ports are set for data input.
Keysight E1458A Command Reference
Range of Values
BYTE none, 0 through 11
WORD none, 0, 2, 4, 6, 8, or 10
LWORd none, 0, 4, or 8
LW64 none or 0
LW96 none or 0
Name of user memory block
(maximum 12 characters)
Default
0
None
93