Ac Power Supply Replacement - Thermo Scientific LevelPRO Series Installation Manual

Continuous gamma level system
Hide thumbs Also See for LevelPRO Series:

Advertisement

AC Power
Replacement
Thermo Fisher Scientific
If this is an AC kit, it will include a new AC power supply that must be
installed on the detector. To do this:
Supply
1. Remove the detector assembly from the housing, unplug the cable at
the bottom of the PCB going to the CPU board and remove the four
screws holding the power supply to the detector chassis (shown in
Figure E–5).
2. Install the new power supply (p/n 886783) onto the detector chassis,
using the four screws removed earlier (shown in Figure E–6).
3. Install the detector back into the housing.
Figure E–5. (left) & Figure E–6. (right)
CE-EMC Protection Installation Procedure

AC Power Supply Replacement

LevelPRO Installation Guide
E-5