3

System Preparation

WARNING:
3.1

Move/Posit the System

Please read and understand the safety precautions before placing the system to ensure
safety for both operator and device.
1. Switch off the power, unplug the power cord.
2. Disconnect all cables from the off-board peripheral devices (printer, recorder, etc.).
3. Release the four casters, and move the system by holding the handle.
4. When you move the system to a desired location, lock the four casters.
5. Leave at least 20cm at the back and both sides of the system.
1. Maintain a generous – free air flowing space around the
CAUTION:
2. Pay extra attention when moving the system on a sloping
1. Do not connect the three-wire cable of the system with
a two-wire plug without protective grounding;
otherwise, electric shock may result.
2. Do connect the power plug of this system to wall
receptacles that meet the ratings indicated on the
rating nameplate. If adapters or multifunctional
receptacles are used, it may cause the leakage current
to exceed the safety requirement.
3. In the environment that patient is 1.5 meters around,
connect peripherals to the auxiliary power outlet
which is capable of isolation protection, or power the
peripherals by auxiliary output cable or isolation
transformer complied with IEC 60601-1-1 or the power
input of the same safety level.
4. DO NOT use power supply of different phases to
power peripherals.
5. When using peripherals not powered by the auxiliary
output of the ultrasound system, or using peripherals
other than permitted by Mindray, make sure the overall
leakage current of peripherals and the ultrasound
system meets the requirement of the local medical
device electrical regulation (like enclosure leakage
current should be no more than 500uA of IEC 60601-1-
1), and the responsibility is held by the user.
back and both sides of the system; failure may result due to
increased rise in system operating temperature.
ground, DO NOT move it on a more than 10°-sloped plane to
avoid system toppling.
System Preparation 3-1