6.3.2
Imaging Display Format Switching When
Frozen
Image display format switching in freeze mode follows the following principles:
2D+M/PW (Press <Freeze> in 2D+M/PW scanning mode)
If the imaging mode before frozen is 2D (frozen) +M/PW (real time) or 2D (real time)
+M/PW (frozen), then in freeze mode, you can switch between 2D (frozen) +PW
(activated) or 2D (activated) +PW (frozen) by pressing <Update> key on the control
panel.
Dual/quad splitting display mode (Press <Freeze> key in dual/quad splitting display
mode)
When enters freeze mode, the default activated window is the real-time window
before frozen. Other image windows display the corresponding cine memories, if
a certain cine memory is empty, then no image is displayed.
Press <Dual> or user defined Quad key to switch between dual-splitting and
quad-splitting modes.
Press <B> button on the control panel to enter the single display format, which
displays the currently activated window. In single display format, press <Dual> or
user defined Quad key to switch between dual-splitting and quad-splitting mode.
Unfrozen: in splitting display status, when you unfreeze the image you can only
unfreeze the image in the activated window, other images still keep frozen. In
single-window display status, the system displays single image after being
unfrozen.
6.4

Cine Review

After you press <Freeze>, the system allows you to review and edit the images prior to
the image frozen. This function is known as cine review. The magnified images can also
be reviewed after <Freeze> is pressed, and the operating method is the same. You can
perform post process operations, measurements, comments adding and body marks on
the images being reviewed.
The system supports manual review as well as automatic review. The default mode is
Manual Cine, but you can switch between Auto Cine and Manual Cine.
In addition, the system supports the images reviewed along with physiological waveforms,
if the detection of physiological waveforms is performed.
CAUTION:
1. Cine Review images can be inadvertently combined in-
between separate patient scans. The cine memory must
be cleared at the end of the current patient and the onset
of the next new patient by selecting the <End Exam> key
on the control panel.
2. Cine files stored in the system's hard drive shall contain
patient information, to avoid the selection of an incorrect
image file and potential misdiagnosis.