Appendix F Batteries - Mindray DC-40 Operator's Manual

Diagnostic ultrasound system
Hide thumbs Also See for DC-40:
Table of Contents

Advertisement

Appendix F Batteries

WARNING:
F.1
Overview
The battery charges when the system is connected to the AC power supply.
Generally, a completely discharged battery will fully charge in less than 6 hours.
When the battery capacity is charged to the full capacity, it can support a standby
period of no less than 24
Note: 1. Charge the battery in time when the battery capacity is empty when necessary;
otherwise, the performance of the batteries is degraded.
2. Power off the system if you will not use the system for a long period of time
(including storage/ transportation condition), and you should not allow the system
in standby status, otherwise the batteries will be out of power and permanently
damaged.
When the external power supply is not connected, two lithium-ion batteries will provide the
power.
Fully charged battery can ensure a normal working condition of more than 30 minutes
(connecting to convex probe when powered on by default without peripherals).
CAUTION:
F.2
Battery Status Indicator
The battery status indicator is located in the lower right corner of the screen, indicating the
battery capacity.
: Indicates the battery capacity is empty.
: Indicates the battery capacity is full.
The battery icons other than the two icons above-mentioned indicate the capacity is
between empty and full.
When the power capacity is not enough, the system will pop up a prompt:
"Warning! Battery is out of power! Please connect to AC power or system will be shut down in
one minute." Connect to AC power to afford normal work.
F.3
One Full Discharge /Charge Cycle
Upon receipt of the system and before first time usage, it is recommended that you perform
one full discharge /charge cycle. If the battery has not been used for over 2 months, you are
recommended to perform one full discharge /charge cycle. It is also recommended that the
battery is stored in the shady and cool area with FCC (full current capacity).
1.
The battery is inside the machine; only technical
professionals from Mindray or engineers authorized
by Mindray after training can perform battery
installation and uninstallation.
2.
If you need to change the battery or buy a new one,
please contact your sales representative.