Mindray DC-40 Operator's Manual page 23

Diagnostic ultrasound system
Hide thumbs Also See for DC-40:
Table of Contents

Advertisement

5.
Turn ON the system only after the power has been turned OFF for a while. If the
system is turned ON immediately after being turned OFF, the system may not be
rebooted properly and could malfunction.
6.
Press <Freeze> key to freeze an image or turn off the power of the system
before connecting or disconnecting a probe. Otherwise, the system and/or
probe can be damaged.
7.
Remove the ultrasound gel from the face of the probe when the examination is
completed. Water in the gel may enter the acoustic lens and adversely affect
the performance and safety of the probe.
8.
You should properly back up the system to a secure external storage media,
including system configuration, settings and patient data. Data stored to the
system's hard drive may be lost due to system failure, improper operation or
accident.
9.
Do not apply external force to the control panel. Otherwise, the system may be
damaged.
10. If the system is used in a small room, the room temperature may rise. Please
provide proper ventilation and free air exchange.
11. To dispose of the system or any part, contact Mindray Customer Service
Department or sales representative. Mindray is not responsible for any system
content or accessories that have been discarded improperly. Mindray is not
responsible for any system content or accessories that have been discarded
improperly.
12. Electrical and mechanical performance may be degraded due to long usage
(such as current leakage or distortion and abrasion); the image sensitivity and
precision may become worse too. To ensure optimal system operations, it is
recommended that you maintain the system under a Mindray service
agreement.
13. The replaceable fuse is inside the chassis. Refer replacing job to Mindray
service engineers or engineers authorized by Mindray only.
14. Do not turn OFF the power supply of the system during printing, file storage or
invoking other system operations. An interrupted process may not be
completed, and can become lost or corrupted.
15. The iScape feature constructs a single extended image from a series of
individual image frames. The quality of the final image is user-dependent and
requires skill to efficiently apply the feature and technique. Exercise caution
when measurements are performed from an iScape image.
16. Ensure that the current exam date and time are the same as the system date
and time.
Please read the following precautions carefully to ensure the safety of the patient and the
operator when using the probes.
1.
WARNING:
2.
3.
The probe is only for use with the specified ultrasonic
diagnostic system. Please refer to the "2.5.2 Probes Available" to
select the proper probe.
The ultrasonic probe must be used only by qualified
professionals.
Confirm that the transducer and probe cable are normal before
and after each examination. A defective probe may cause
electric shock to the patient.
Safety Precautions 1-7