Mindray DC-40 Operator's Manual page 95

Diagnostic ultrasound system
Hide thumbs Also See for DC-40:
Table of Contents

Advertisement

Increasing the gain will increase the flow signal presented and noise. The
Effects
signals may be missing when the gain is adjusted too low.
Map
Description This feature indicates the display effect of the Power image.
The maps in the Power mode image are grouped into two categories:
Power maps and Directional Power maps.
To select from among the maps, turn the knob under [Map] on the touch
Operations
screen.
There are 8 kinds of maps provided: P0-3 belong to Power mode maps,
while dP0-dP3 belong to Directional Power mode maps.
The Power maps provide information about blood flow, which are highly
sensitive to the low-velocity flows.
The Directional Power maps provide information about flow direction.
Dynamic Range
Description This function adjusts the transformation of echo intensity into color signal.
Adjust through the [Dyn Ra.] item on the touch screen to adjust the
Operations
dynamic range.
The adjusting range is 10-70dB in increments of 5dB.
Increasing the dynamic range leads to higher sensitivity to low-power
Effects
signals, thus enhancing the range of signals to display.
Image Optimization 5-19