Limits Of Acoustic Output - Mindray DC-40 Operator's Manual

Diagnostic ultrasound system
Hide thumbs Also See for DC-40:
Table of Contents

Advertisement

adjusts the acoustic output as measured in a water tank to account for the effect of
ultrasound propagation through tissue. By convention, a specific average intensity
attenuation value is used, which corresponds to a loss of 0.3 dB/cm/MHz. That is, the
intensity of ultrasound will be reduced by 0.3 dB/MHz for every centimeter of travel from
the transducer. This can be expressed by the following equation:
Where I
is the attenuated intensity, I
atten
distance z), fc is the center frequency of the ultrasound wave (as measured in water),
and z is the distance from the transducer. The equation for attenuating pressure values is
similar except that the attenuation coefficient is 0.15 dB/cm/MHz, or one-half the intensity
coefficient. The intensity coefficient is double the pressure coefficient because intensity is
proportional to the square of pressure.
Although the attenuation coefficient chosen, 0.3 dB/cm/MHz, is significantly lower than
any specific solid tissue in the body, this value was chosen to account for fetal
examinations. In early trimester ultrasound fetal examinations, there may be a significant
fluid path between the transducer and the fetus, and the attenuation of fluid is very small.
Therefore the attenuation coefficient was lowered to account for this case.

14.7.2 Limits of Acoustic Output

In accordance with the FDA Track 3 requirements, the derating (or attenuated) approach
was incorporated into the FDA Acoustic Output Limits, as listed below. The maximum
acoustic output level from any transducer in any operating mode is expected to fall below
these limits.
FDA Maximum Acoustic Output Limits for Track 3 (Attenuated Values)
Application
Regions
(except
eyes)
14.7.3 Differences between Actual and Displayed MI
and TI
In operation, the system will display to the operator the Acoustic Output Parameters
Thermal Index, TI, or Mechanical Index, MI (or sometimes both parameters
simultaneously). These parameters were developed as general indicators of risk from
either thermal or mechanical action of the ultrasound wave. They serve to indicate to the
operator whether a particular setting of the system increases or decreases the possibility
of Thermal or Mechanical effect. More specifically, they were designed to assist in the
implementation of the ALARA principle. As an operator changes a given system control,
the potential effect of the change in output will be indicated. However, the Thermal Index
is not the same as temperature rise in the body, for several reasons. First of all, in order
to provide a single display index to you, a number of simplifying assumptions had to be
made. The biggest assumption was the use of the attenuating formula described above,
which is much lower than the actual value for most tissues within the body. Scanning
through muscle or organ tissue, for example, will produce much higher attenuation than
0.3 dB/cm/MHz. There were also significant simplifications made for the thermal
=
×
I
I
10
atten
water
is the intensity measured in a water tank (at
water
I
I
spta.3
sppa.3
2
(mW/cm
)
(W/cm
720
≤ 190
×
×
(
. 0
/ 3
10
f
z
)
c
2
)
or
MI
≤ 1.9
Acoustic Output 14-5