TABLE 3
GUIDANCE AND MINDRAY DECLARATION—ELECTROMAGNETIC IMMUNITY
The system is intended for use in the electromagnetic environment specified below. The customer
or the user of system should assure that it is used in such an environment.
IMMUNITY TEST
Conduced RF
IEC 61000-4-6
Radiated RF IEC
61000-4-3
Note 1 At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and human body.
Field strengths from fixed transmitters, such as base stations for radio (cellular /cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic
site survey should be considered. If the measured field strength in the location in which
system is used exceeds the applicable RF compliance level above, system should be
observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating system.
Over the frequency ranges 150kHz to 80MHz, field strengths should be less than 3V/m.
15-4 EMC Guidance and Manufacturer's Declaration
IEC 60601
COMPLIANCE
TEST LEVEL
3 Vrms
150 kHz - 80
MHz
3 V/m
80MHz -
2.5GHz
LEVEL
ENVIRONMENT-GUIDANCE
Portable and mobile RF
communications equipment should be
used no closer to any part of system,
including cables, than the
recommended separation distance
calculated from the equation applicable
3 Vrms
to the frequency of the transmitter.
Recommended separation distance
d = 1.2 x
d = 1.2 x
d = 2.3 x
Where, P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation distance in
meters (m).
3 V/m
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, should be
less than the compliance level in each
frequency range.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
ELECTROMAGNETIC
P
80 MHz to 800 MHz
P
800 MHz to 2.5GHz
P