13.2.4 Verifying the Biopsy Guide Line

Adjusting the needle mark is necessary before every biopsy procedure.
1. Confirm that the needle-guided bracket has been installed securely in the correct
position.
2. Prepare a container filled with sterile water.
3. Place the head of the probe in the sterile water, and get a biopsy needle into the
needle guide.
4. When the biopsy needle appears on the image, confirm that the biopsy needle is
displayed at almost the same position as the selected needle mark.
WARNING:
You can perform guide line verification on a single live B/C image, and all
NOTE:
biopsy-irrelevant operations are forbidden.
Ensure that all guide parts are seated properly prior to
performing a biopsy.
1. Prior to each biopsy procedure, be sure to verify the
guide line.
2. If the needle is not consistent with the guide line, DO
NOT perform the biopsy procedure.