NGB-011
Locating
Clamp
pit
Locating groove
Needle-guided
bracket
Probe
13.2.3 Needle-guided Bracket Inspection and
Installation
Inspection of the Needle-guided Bracket
Be sure to perform inspections before and after use of the needle-guided bracket. If an
abnormality is found on the needle-guided bracket, immediately stop using it and contact
MINDRAY Customer Service Department or sales representative.
1. Sterilize the needle-guided bracket before and after use.
2. Confirm that the needle-guided bracket is free of damage, deformation, stripping,
and malfunction, loose or missing parts.
3. Confirm that the needle-guided bracket is securely mounted in the correct position.
Installing the Needle-guided Bracket
NGB-004
(1) Put on the sterile probe sheath.
Needle guide
Clamping knob of the
needle guide
Grip
Needle guide rack
knob
Needle guide
hole
Probes and Biopsy 13-19