Navigating the Screens
and then the change is confirmed by pressing Accept. In either case, press the Accept button after
each setting change or make multiple changes and then press Accept.
If you decide not to make the changes you started, press the Cancel button instead of pressing
Accept or just wait and the values will revert back to the original settings.
The mode and mandatory breath type selections determine which Main screen breath delivery
parameters are active and available for adjustment. Follow these steps when setting up ventila-
tion:
While the Main screen is in view, start by selecting the mode and mandatory breath type along the left
1.
margin of the screen.
Adjust all Main screen parameters that are visible.
2.
Press the Accept button to implement the mode/breath type and relevant parameters change.
3.
Visit the More screen to turn NIV on or off.
4.
When NIV is turned on while PEEP is in use, the bias flow adjustment window appears so that bias flow
5.
can also be adjusted.
Adjust the other relevant ventilation parameters on the More screen.
6.
Visit the Alarm screen to adjust/check alarm parameters.
7.

Startup Screen Navigation (Standby Condition Only)

4.4
When the ventilator is turned on, it goes through a short self-test before entering standby condi-
tion. During the short self-test, confirm the alarm sounds and the LEDs light.
While in standby, the Startup screen is available. In standby, ventilator settings can be adjusted.
The Startup screen includes three buttons in the top message area: Start Ventilation, Circuit Check,
and Activate Presets.
4-4
Figure 4-2.  Startup Screen (Standby condition)
Operator's Manual