5.3.3 "Alarm Record" screen
The screen displays the alarm record occurred in the past. The ten newest errors are logged. The oldest error
will be deleted one by one every time a new alarm occurs. New errors are displayed from the top on the
screen.
It displays the date, time and component when an alarm occurred, alarm contents, recovery date, and recov-
ery time from the left to right.
If you turn off the power in a state that the alarm is generated, the highlighted time and date when the power
was turned off are displayed for "Alarm recovery time" and "Alarm recovery date".
All the alarms except "Connection Error" which indicates communication failure between the receiver unit
and the control unit will be activated after at least five minutes from the power off. The alarm of "Low Light
Transmission" is activated after 1 minute continuation. Though it takes 6 minutes from just after the power
on.
5.3.3.1 Name (functions)
(1) Alarm occurrence date ············ Displays the date when device failure, high gas temperature or
(2) Alarm occurrence time ············ Displays the time when alarm occurred.
(3) Alarm occurrence component ···· Displays the component and analog input for which alarm oc-
(4) Alarm contents ····················· Displays the contents of alarm.
(5) Alarm recovery date ··············· Displays the date when alarm is recovered. Nothing is displayed
(6) Alarm recovery time ·············· Displays the time when alarm is recovered. Nothing is displayed
(7) Select Delete key ·················· Deletes the selected alarm.
(8) All Delete key ······················ Deletes all alarms.
alarm occurred.
curred.
for the alarm which is not recovered.
for the alarm which is not recovered.
- 36 -