6.11.3 Operation method
(1) Remove the cap from the gas inlet and the outlet of the check cell. Connect the inlet side to the N
cylinder.
(2) Feed N
gas to the check cell to obtain the concentration of the target gas inside the stack.
2
(3) Feed the gas, which was obtained by "6.11.2 Concentration of the gas which is fed to the check cell and
gas concentration fluctuation", to the check cell.
(4) After the indication value become stable, check the difference from the span point.
(5) Fill the check cell with N
(6) Put on the caps so that the measured gas component does not come into the check cell.
INZ-TN8ZSS-E
gas.
2