Fuji Electric ZSS-8 Series Instruction Manual page 77

Cross stack laser gas analyzer
Table of Contents

Advertisement

6.5.3 Moving average time
This parameter is required if you set the indication and the output to the average in "6.6.1 Instantane-
ous/average value". Setting range is from 1 to 60 minutes. Up to the time you set, the analyzer emit the av-
erage value at the time.
The average value is reset when you press the ENT key after changing the setpoint.
Averaging operation
When the moving average time is set to 60 minutes
0
Sampling rate: twice per second
Mode 1:
Until the time reaches 60 minutes, the analyzer calculates the average by dividing the available values
by the time passed.
*Available values: the values obtained after you reset the average to the each moment.
The quation is:
Mode 2:
Until the time reaches 60 minutes, the analyzer calculates the average by dividing the available values
by 60 minutes.
The equation is:
In both modes, it is after 60 minutes that you can obtain the first moving average value for 60 minutes.
INZ-TN8ZSS-E
Reset
60min
Σ
Σ
60
- 67 -