4.1.2 Control unit
Select a location that meets the following conditions.
(1)
Ambient temperature : –5 to 45°C
(2)
Ambient humidity
(3)
Power supply
(4)
A place where the instrument receives no vibration
(5)
A place where the ambient air is clean
(6)
A place where has enough space for maintenance work or inspection
(7)
A place where induced electrical noise, such as large electrical current or spark, is low
(8)
Install a breaker that meets the requirements of IEC60947-1 and IEC60947-3 in the facility in which
the analyzer is placed.
Notes:
• The breaker should be installed close enough to the analyzer so that an operator can use it without dif-
ficulty.
• The breaker should be identified as the one for this product.
• The breaker must meet the rating of the analyzer.
: 90 % RH or less
: Rated voltage
: 100V to 240V AC
Rated frequency : 50Hz/60Hz
- 10 -
INZ-TN8ZSS-E