(5) Attach two Rc1/4" pipe fittings to the inlet and outlet of the calibration gas cell. Connect the tube from
the standard gas (N
(6) Connect the exhaust gas tube to the outlet side. For exhaust gas line, use a tube as large and short as
possible.
*If you use thermometer and/or pressure gauge, attach them on the exhaust gas line, and connect their
4–20mA output to the analog input terminal of the control unit. Make settings for the analog input in
reference to 6.7.1 "Analog input setting: sensor input".
(7) After all the pipes and cables are connected, flow the calibration gas with a flow rate of 1.5 to 2.0
L/min.
) cylinder to the inlet side.