2.7.3.

Operation with an External Sensor (External Input)

In some instances, it might be desirable to use an external sensor to provide process information to the
control circuitry in a Hastings flow instrument. For example, you might want to control the pressure in a
vacuum system by adjusting the rate at which the system is backfilled with a gas. Hastings flow
controllers have provision for accepting a 0–5 VDC output from an external sensor at pin 13 of the DA-15
(15 Volt), Pin 1 of the DE-9 (24 Volt) connector.
2.7.3.1.

15 Volt Units

To activate this feature, the cover of the
HFC must be removed to gain access to the
electronics card and move a jumper. For
the 15 Volt units, J6 is located on the left
side of the pcb between the span
potentiometer (R7) and the large D
connector (J1). In the normal operating
mode, the jumper covers the left two pins.
To select an external sensor input, move
the jumper to the right two pins. This
swaps the flow input to the controller
circuit from the flow meter output to pin
13 of the "DA-15" connector.
2.7.3.2.
24 Volt Units
For the 24 Volt units J8 is a six-pin jumper
block located down and to the right from
the DC-DC converter. There are three
jumpers installed horizontally. The middle
jumper ties the flow output to the valve
control circuit and the top jumper ties Pin
1 of the DE-9 connector to reference
voltage circuit. Remove both of the upper
two jumpers and reinstall one jumper
vertically between the top two pins on the
right side (see diagram below). This will tie
Pin 1 of the DE-9 connector to the valve
control circuit and allow the controller to
maintain the external process variable
equal to the desired setpoint. The 5 Volt
reference is not available in this
configuration.
EAR99 Technology Subject to Restrictions Contained on the Cover Page