Step 1. Connect the equipment as shown in Figure 2-7 on page 40.
Step 2. Set the Synthesized Sweeper as follows:
Test Descriptions
Frequency Accuracy

Test Setup

Frequency Accuracy test setup
Ref In
Ref In
Input
RF Cable

Test Procedure

Throughout the Frequency Accuracy Test Procedure the term 'the
Worksheet' refers to the Frequency Accuracy Test Worksheet on page 45
and the term 'the Test Record' refers to the Frequency Accuracy Test
Procedure Test Record detailed in the relevant Appendix for the model
number being tested.
1. Press the
Preset
2. Set the power level to -70dBm (press the
).
dB(m)
3. Switch
RF ON
40
BNC Cable
SYNTHESIZED
SWEEPER
10 MHz
RF Output
Adapter 2
key and wait for the preset routine to complete.
FREQUENCY
STANDARD
key and
Power Level