Figure 2-20
Table 2-21
Equipment description
Noise source
Noise source
Test Descriptions
Instrument Noise Figure
Typical plots with limit lines for the N8975A
Related Adjustment
None
Required Test Equipment
Required equipment for Instrument Noise Figure Test
Critical specification for equipment
substitution
Frequency Range: 10 MHz to 3 GHz
Typical ENR: 4.5 - 6.5 dB
Connection: 3.5 mm (M) or N Type
Frequency Range: 10 MHz to
26.5 GHz
Typical ENR: 12.0 - 17.0 dB
Connection: 3.5 mm (M) or N Type
76
Recommended model
346A standard or Option
001
346C
Chapter 2