Figure 2-5
Step 9. As the calibration kit selection is dependant on the NFA being tested,
NOTE
Step 10. Select an S11 port calibration (press the
Step 11. Connect the Open to the test port or, if fitted, the adapter and perform
Step 12. Connect the Short to the test port or, if fitted, the adapter and perform
Step 13. Connect the Load to the test port or, if fitted, the adapter and perform
Step 14. Save the calibration (press the
Test Descriptions
Input VSWR
Vector Network Analyzer: Adaptor Connection
VECTOR NETWORK
ANALYZER 2
Port 1
select the relevant calibration kit in use from the Cal menu (press the
key,
selection and
Calibration Kit
The connector type applies to the test port or adaptor if fitted and not the
NFA under test.
).
Reflection 1-port
the open calibration (press
key).
the open calibration (press
key).
the open calibration (press
).
done 1-port CAL
34
uWave Cable
).
Return
and then select the relevant menu
Opens
and then select the relevant menu
Shorts
, select the relevant menu key
Load
Save/recall
Adapter 3
key, then
Cal
Calibrate Menu
key and then
Save State
Cal
and
and then
).
Chapter 2