Keysight NFA Series Performance Verification And Calibration Manual page 111

Noise figure analyzers
Hide thumbs Also See for NFA Series:
Table of Contents

Advertisement

Tuning Accuracy
(Start, Stop,
Centre, Marker)
Frequency (MHz)
At measurement bandwidth of 4 MHz
10 - 3000
> 3000 - 26500
At measurement bandwidth of < 4 MHz
10 - 3000
> 3000 - 26500
Chapter 3
Technical Specifications
o
Temperature (0
C - 55
± Reference error + 100 kHz
± Reference error + 400 kHz
± Reference error + 20 kHz
± Reference error + 20% of measurement
bandwidth
Frequency
o
C)
91