Keysight NFA Series Performance Verification And Calibration Manual page 51

Noise figure analyzers
Hide thumbs Also See for NFA Series:
Table of Contents

Advertisement

Measuring VSWR from 1000 MHz to 1500 MHz
Step 1. Connect the test equipment as shown in Figure 2-4 on page 28.
Step 2. Set the Network Analyzer Start/Stop frequencies from 1000 MHz to
1500 MHz.
Press the
Start
G/n
Step 3. Set the fixed frequency of the NFA to 1250 MHz (press the
Frequency/Points
Step 4. Restart the measurement sweep on the DUT (press the
Step 5. Using the marker peak function on the network analyzer, determine the
maximum displayed VSWR over the measured frequency range (press
the
Marker Fct
Step 6. Record the maximum displayed signal frequency and VSWR on the Test
Record.
Step 7. Verify that the measured VSWR from 1000 MHz to 1500 MHz is within
the published specification.
Chapter 2
key and then
,
,
1
.
0, G/n;
key and then
Fixed Freq
n key,
Mkr Search [ON]
then press the
Stop
,
,
).
1, 2, 5, 0
MHz
then
).
Search: Max
Test Descriptions
Input VSWR
key and
,
1, ., 5
key).
Restart
31

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

N8972aN8973aN8975aN8974a

Table of Contents