Table 2-12
Equipment description
Synthesized Sweeper
Attenuator/Switch Driver
1 dB Step Attenuator
Adapter 1
Adapter
RF Cable (X 2)
Test Descriptions
Noise Figure Range and Accuracy
Related Adjustment
None
Required Test Equipment
Noise Figure Range and Accuracy test equipment
Critical specification for equipment
substitution
Frequency Range: 50 MHz
Frequency Accuracy (CW): 0.02%
Power Level Range: -54dBm to -65dBm
Compatible with 8494G/H Programmable step
attenuators
Attenuation Range: 0 to 11dB
Frequency Range: 50 MHz
Connectors: 3.5mm (F)
Calibrated at 50 MHz with accuracy of 0.010 dB
N8972/3A N Type (M) to 3.5mm (F)
N8974/5A 3.5mm Precision (F) to 3.5mm (F)
83620/30 3.5mm Precision (F) to 3.5mm (F)
83640/50 2.4mm Precision (F) to 3.5mm (F)
50 MHz 3.5mm (M) to 3.5mm (M), 61cm (24 in.)
56
Recommended model
83620/30/40/50B
Option 001 and 008
11713A
8494G/H Option 002
1250-1744
1250-1749 or 83059B
1250-1749 or 83059B
11901B
11500E