Keysight NFA Series Performance Verification And Calibration Manual page 71

Noise figure analyzers
Hide thumbs Also See for NFA Series:
Table of Contents

Advertisement

Figure 2-12
Step 2. Calculate the half power level from the data as shown in Figure 2-12.
Step 3. Use the summed value column as shown in Figure 2-12 as a look up
Example 2-2 summed P
Equation: ½ Power = Summed P
reference and find the cell, which closest matches the ½ Power Reading.
The corresponding frequency column is given as the center frequency
point. Enter the reading into the Test Worksheet Table on page 45 and
ensure the measured value is within its published specification.
Chapter 2
for reading 1 to reading 6
hot
Reading 201 / 2
hot
Test Descriptions
Frequency Accuracy
51

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

N8972aN8973aN8975aN8974a

Table of Contents