Figure 2-13
NOTE
Step 1. Connect the equipment as shown in Figure 2-13.
Step 2. Select the DC range on the digital multimeter (press
Step 3. Ensure that the preset is set to factory settings (press the
Step 4. On the NFA press the
Step 5. Set sweep to single (press
Step 6. Set the noise source off (press the
Step 7. Enter the digital multimeter reading into the Test Record as the Noise
Test Setup
Equipment setup for Noise Source Supply Accuracy test
DIGITAL
MULTIMETER
Adapter on
(2 Wire) HI/LO
Test Procedure
Throughout the Noise Source Supply Accuracy Test Procedure the term
'the Test Record' refers to the Noise Source Supply Accuracy Test Record
detailed in the relevant Appendix for the model number being tested.
menu).
FUNCTION/RANGE
then
,
More
Power On (Preset)
. Press
Service
Enter Password
Press
then
Enter
Service
Source Supply Off measurement.
Chapter 2
Noise Source Supply Accuracy
+28V Noise
Source Supply
BNC Cable
and
Preset (Factory)
key and wait for the routine to finish.
Preset
,
Sweep
Sweep Mode: single
key, then
System
then enter the service password -2010.
and
Noise Source Off
Test Descriptions
NOISE
FIGURE
ANALYZER
in the
DCV
System
).
).
,
and
More
More
).
key,
53