Measurement Jitter - Keysight NFA Series Performance Verification And Calibration Manual

Noise figure analyzers
Hide thumbs Also See for NFA Series:
Table of Contents

Advertisement

Equation 2-2

Measurement Jitter

Test Description
A noise source is connected to the NFA's input. The Analyzer then
measures linear Y-Factor over 100 samples. The standard deviation is
then calculated using the "non-biased" or "n-1" method. STDEV uses the
following formula:
n
σ
=
---------------------------------------- -
The linear value is then converted to Log10 to give the Y-factor Standard
Deviation with no averaging.
Test Specification
Jitter with no averaging: Y-factor Standard Deviation <0.10dB
Related Adjustment
None
Chapter 2
2
2
x
x
(
)
n n 1
Test Descriptions
Measurement Jitter