Before You Start - Keysight NFA Series Performance Verification And Calibration Manual

Noise figure analyzers
Hide thumbs Also See for NFA Series:
Table of Contents

Advertisement

CAUTION
CAUTION
Preparing for Calibration and Performance Verification Test

Before You Start

Before You Start
Switch on the Noise Figure Analyzer and let it warm up for 1 hour.
Read the rest of this section before you start any of the tests, and make a
copy of the relevant test records for the model you are testing provided in
Appendices A through D.
Ensure that the recommended torque settings detailed in Table 1-3 are
adhered to at all times.
Ensure that a valid ENR measurement table has been loaded from the C:
or A: drive before carrying out any performance test. The analyzer will
not perform a measurement without an ENR table loaded.
4
Chapter 1

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

N8972aN8973aN8975aN8974a

Table of Contents