Keysight NFA Series Performance Verification And Calibration Manual page 107

Noise figure analyzers
Hide thumbs Also See for NFA Series:
Table of Contents

Advertisement

Step 9. Convert the linear standard deviation to Log
Standard Deviation
Table 2-24
Y-Factor
Y-Factor
Y-Factor
Reading
Reading
Reading
(Lin)
(Lin)
(Lin)
1-10
11-20
21-30
10 * LOG
(1+ standard deviation
10
to give the standard deviation for Y-Factor results with no Averaging.
Enter the result into the Test Record ensuring that the measured value
is within its published specification.
Example:
=
10 * LOG
=
10 * LOG
=
0.0478dB
Measurement Jitter Test Worksheet
Y-Factor
Y-Factor
Reading
Reading
(Lin)
(Lin)
31-40
41-50
Chapter 2
)
Lin
(1+ standard deviation
10
(1+ 0. 011071
)
10
Lin
Y-Factor
Y-Factor
Reading
Reading
(Lin)
(Lin)
51-60
61-70
Test Descriptions
Measurement Jitter
using
10
)
Lin
Y-Factor
Y-Factor
Y-Factor
Reading
Reading
Reading
(Lin)
(Lin)
(Lin)
71-80
81-90
91-100
87

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

N8972aN8973aN8975aN8974a

Table of Contents