Figure 2-2
NOISE FIGURE
ANALYZER
NOTE
NOTE
Test Setup
Before starting the 10 MHz Out Frequency Reference Adjustment,
connect the equipment as shown in Figure 2-2.
10 MHz Out Frequency Reference Adjustment Test Setup
10 MHz
Ref Out
UNIVERSAL
COUNTER
BNC Cable
Test Procedure
Before performing any adjustments ensure that the 10 MHz Frequency
Reference Accuracy test has been done and that the measured frequency
is outside the published specification.
The NFA must be on and in internal frequency mode for at least 1 hour
before you start the test.
Chapter 2
10 MHz Out Frequency Reference Adjustment
BNC Cable
10 MHz
Ref In
Channel 1
Test Descriptions
FREQUENCY
STANDARD
19