Keysight NFA Series Performance Verification And Calibration Manual page 27

Noise figure analyzers
Hide thumbs Also See for NFA Series:
Table of Contents

Advertisement

Table 1-1
Equipment description
3.5mm Calibration Kit
Synthesized Sweeper
Digital Multimeter
Attenuator/Switch Driver
1 dB Step Attenuator
Noise Source
Noise Source
Preparing for Calibration and Performance Verification Test
Recommended Test Equipment
Critical specification for equipment
substitution
Impedance: 50
Frequency Band: 3 GHz to 26.5 GHz
UUT: N8974A and N8975A
Frequency Range: 10 MHz to 26.5 GHz
Frequency Accuracy (CW): 0.02%
Power Level Range: -55 dBm
Input Resistance of 10 MΩ
Accuracy of 10 mV on 100 V range
Compatible with 8494G/H Programmable
step attenuators
Attenuation Range: 0 to 11 dB
Frequency Range: 50 MHz
Connectors: 3.5 mm (F)
Calibrated at 50 MHz with accuracy of
0.010 dB
Frequency Range:10 MHz to 6.7 GHz
Typical ENR: 4.5 - 6.5 dB
Connection: 3.5 mm(M) or Type N
Frequency Range: 10 MHz to 26.5 GHz
Typical ENR: 12.0 - 17.0 dB
Connection: 3.5 mm(M)
Chapter 1
Recommended Test Equipment
Recommended
model
85052D
83620/30/40/50B
Option 001 or 008
3458A
11713A
8494G/H Option 2
346A Std or Option
001
346C
7