Test Descriptions
Noise Source Supply Accuracy
Step 8. Set the noise source on (press the
. Press
Service
service password -2010. Press
Step 9. Enter the digital multimeter reading in the Test Record as the Noise
Source Supply On measurement.
Step 10. Press the
Step 11. Verify that the measured values are within the published specifications.
54
Noise Source On
key to return to the default Noise Source Supply State.
Preset
key, then
System
). Press
Enter Password
then
and
Enter
Service
,
and
More
More
then enter the
).
Noise Source On
Chapter 2