Keysight NFA Series Performance Verification And Calibration Manual page 55

Noise figure analyzers
Hide thumbs Also See for NFA Series:
Table of Contents

Advertisement

Figure 2-6
Step 1. Connect the test equipment as shown in Figure 2-6 on page 35.
Step 2. Set the Network Analyzer Start/Stop to measure the DUT's input VSWR
Step 3. Set the fixed frequency of the NFA to 5000 MHz (press the
Step 4. Restart the measurement sweep on the DUT (press the
Step 5. Using the marker peak function on the network analyzer, determine the
Step 6. Record the maximum displayed signal frequency and VSWR on the Test
Step 7. Verify that the measured VSWR from 3000 MHz to 6700 MHz is within
Measuring Input VSWR from 3000 MHz to 6700 MHz
Input VSWR Test Setup 2
VECTOR NETWORK
ANALYZER 2
Port 1
uWave Cable
from 3000 MHz to 6700 MHz.
Press the
key and then
Start
G/n
key and then
Frequency/Points
maximum displayed VSWR over the measured frequency range (press
the
n key,
Marker Fct
Record.
the published specification.
Chapter 2
NOISE FIGURE
ANALYZER
Input
,
then press the
3,., 0
G/n;
Fixed Freq
then
Mkr Search [ON]
Test Descriptions
Adapter 3
key and
Stop
,
).
5, GHz
Restart
).
Search: Max
Input VSWR
,
,
,
6
.
7
key).
35

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

N8972aN8973aN8975aN8974a

Table of Contents