Table 2-22
Equipment description
Noise source
Adapter
BNC cable
Figure 2-26
N8972A
N8973A
Type N
Opt 001
OR
N8974A
N8975A
3.5mm
Std.
Test Descriptions
Measurement Jitter
Required Test Equipment
Required equipment for Instrument Measurement Jitter Test
Critical specification for equipment
substitution
Frequency Range: 1 GHz
Typical ENR: 4 - 7dB
Connection: 3.5 mm(M) or N Type(M)
3.5 mm Precision (F) to 3.5 mm(F)
Length 122 cm (48 in.), Frequency
10 MHz
Test Setup
Measurement Jitter Test Setup
346A
NOISE FIGURE
ANALYZER
Adapter
346A
Input
NOISE
SOURCE
84
Load ENR Data from C:
+28V Noise
Source Supply
Create File from Data on
BNC Cable
Noise Source Label
Recommended model
346A Standard or option
001
83059B
10503A
Insert ENR Disk
and Load from A:
OR
OR